0

No products in the cart.

GDP-C Gas Permeability Tester | Brugger Feinmechanik

GDP-C Gas Permeability Tester is designed for testing the permeability of films. It supports an extended temperature range through external water bath circulator.


Request Quote

Please select a valid form

Description

GDP-C Gas Permeability Tester | Brugger Feinmechanik

Brugger Feinmechanik’s GDP-C is for testing the permeability of films. It supports an extended temperature range through external water bath circulator.

Additional information

Test Method

Quantitative determination of the gas permeability of film materials using differential pressure (manometric method)

Test Gas

All non-corrosive and non-flamable gases

Measuring Range

0.5 cm³/(m²×d×bar) to 30,000 cm³/(m²×d×bar)

Resolution

0.1 cm³/(m²×d×bar)

Main Pressure Sensor (MPS)

piezoresistive element, temperature stabilised

MPS Measuring Range

0 hPa to 1100 hPa

MPS Tolerance (absolute)

< 0.3 % of the final pressure (1000 hPa)

MPS Resolution

0.01 hPa (display) , 21 Bit (internal)

MPS Data Storage

up to 4096 test values (with dynamic time interval selection)

Pressure Absorption Gauge (PAG)

piezoresistive element (0 hPa to 2000 hPa)

PAG Tolerance

< 1% of the final pressure (1000 hPa)

PAG Resolution (Display)

0.1 hPa

PAG Data Storage

up to 512 test values (width dynamic time interval selection)

Sample Temperature Range (TR)

-20 °C to 80°C (with external thermostat)

TR Tolerance (temperature sensors)

+/- 1.5 °C

TR Resolution of the temperature display

0.1 °C

TR Data Storage

up to 512 test values (with dynamic time interval selection)

PC Interface

Serial interface RS232, 115.200 Baud

Dimensions

50 x 31 x 74 cm

Weight

20 kg

Storage Temperature

0 °C – 50 °C

Working Temperature

20 °C – 26 °C (to be constant during operation when testing films with a low GTR)

Relative Humidity

max. 80 %, non-condensing

Electrical connection GDPC

230 V / 50 – 60 Hz, power consumption ca. 50 W, approx.

Electrical connection PC

230 V / 50 – 60 Hz, power consumption ca. 50 W, approx.

Electrical connection monitor

230 V / 50 – 60 Hz, power consumption ca. 150 W, approx.

Electrical connection(Thermostat)

230 V / 50 – 60 Hz, power consumption ca. 2300 W, approx.

Standard

ISO 15105-1 / ASTM D1434-82(2015) procedure M