Description
GTT Gas Transmission Tester | Brugger Feinmechanik
Brugger Feinmechanik’s GTT is for the determination of the gas permeability of films. It has an integrated touch panel for control and evaluation, as well as an intern temperature control.
- Different testing gases
- Integrated panel PC
- Internal temperature control
- Enhance evaluation
How it works
GTT is a gas permeability tester that combines cutting-edge technology with the proven robust measuring principle of its predecessor. (GDP-C).
The GTT is an exceptionally compact device. It includes a high-resolution touchscreen for inputting all data needed for measurements and outputting analysis results.
Data entry available via wireless keyboard (and wireless mouse). Even more connectivity options are provided in the form of two readily available USB connections and a LAN port. There is no need for an additional PC to control the GTT or evaluate its data.
This new model has replaced the extremely energy-hungry external water bath thermostat with an energy-efficient electrothermal temperature control device. This means that the GTT is easy to install without extra apparatus (except a vacuum pump).
The GTT also boasts the following features
- The device includes no wear parts, making it exceptionally low-maintenance.
- Includes a highly robust sensor (piezoresistive pressure sensor) and can withstand the sudden ingress of air during a measurement. It is also immune from the oversaturation that can affect electrochemical sensors.
- The measurement principle used means that the requirements for generating a vacuum are relatively low. As the starting pressure level has no impact within a broad range (up to 10 hPa), all that is needed is a simple rotary vane pump.
- Directly measures gas transmission of a wide range of test gases increasing the number of uses for this device. For instance, the properties of the materials being tested can often be measured much faster and more precisely. The device does not require any modifications, let alone any intervention, to be used for different test gases.
- The device comes with a gas-saving mode in which the flow rate of the test gas can be reduced to approximately 1.2 l per hour. For example, the contents of a small 10 l bottle can give 60 days of uninterrupted measurement.
- Several measurement modes enable values including the time lag to be calculated. This is needed in order to determine further physical properties such as the solubility or the diffusion coefficient of the sample.
- The temperature is controlled electrothermally by ultra-precise, energy-efficient Peltier elements during the sampling process. This covers the full temperature range of 15 °C to 45 °C that is conventionally used for measurements. This gives the user numerous options for measuring other material properties, such as the
activation energy. - Potential uses for the GTT: humidified test gases (digital sensor located right inside the top sample chamber). This relays continuous measurements of the relative humidity and temperature of the test gas.